Technology
Laser package
Sensitivity
Accuracy / repeatability
Calibration
Analysis range
PML2 sample volume
Operation
Viscosity range
Operating temperature
Fluid compatibility
Typical test time
Power
Data storage
Computer interface
Hose connections
Dimensions
Optional product |
Automatic optical particle analyser
Twin laser and optical diode detectors
>4,6,14,21,25,38,50,68µm(c), micron range revised ISO 4406 Standard
Better than 3% typical
Each units is individually calibrated with ISO Medium Test Dust (MTD) as based on ISI 11171:1999
ISO 8 to ISO 24, ISO 4406 Code. (NAS 1638 - 2 to 12)
(SAE AS 4059 - 2 to 12)
15ml
System working pressure - Maximum 400 bar, Minimum 2 bar (Series 30)
10 bar (Series 31)
To 400 centistrokes
+5 to +80°C
Mineral oil and petroleum based fluids, and Skydrol® (consult MP Filtri on other fluids
2 minutes
12/24 volt DC power supply
600 tests
RS 232 communication port
Microbore pressure hose 1500mm long with minimess fittings (5m and 10m lengths available)
Quick coupling waste hoses
Height 120mm, Length 275mm, Width 250mm, Weight 4.8 kilos
In-line coarse screen unit minimess fitting
500 micron stainless steel cleanable mesh 400 bar filter pressure
|