Bottle sampling options
Accessories
Features + benefits
Re-calibration
Software
Applications, testimonials
LPA2 Ordering information
Manuals + brochures
Contact

See HOW TO ORDER below the PML - PML2-W technical specification

 

PML2 - PML2-W SPECIFICATION

 

Technology
Laser package
Sensitivity
Accuracy / repeatability
Calibration

Analysis range

PML2 sample volume
Operation

Viscosity range
Operating temperature
Fluid compatibility

Typical test time
Power
Data storage
Computer interface
Hose connections


Dimensions
Optional product

Automatic optical particle analyser
Twin laser and optical diode detectors
>4,6,14,21,25,38,50,68µm(c), micron range revised ISO 4406 Standard
Better than 3% typical
Each units is individually calibrated with ISO Medium Test Dust (MTD) as based on ISI 11171:1999
ISO 8 to ISO 24, ISO 4406 Code. (NAS 1638 - 2 to 12)
(SAE AS 4059 - 2 to 12)
15ml
System working pressure - Maximum 400 bar, Minimum 2 bar (Series 30)
10 bar (Series 31)
To 400 centistrokes
+5 to +80°C
Mineral oil and petroleum based fluids, and Skydrol® (consult MP Filtri on other fluids
2 minutes
12/24 volt DC power supply
600 tests
RS 232 communication port
Microbore pressure hose 1500mm long with minimess fittings (5m and 10m lengths available)
Quick coupling waste hoses
Height 120mm, Length 275mm, Width 250mm, Weight 4.8 kilos
In-line coarse screen unit minimess fitting
500 micron stainless steel cleanable mesh 400 bar filter pressure

HOW TO ORDER

 

< select buttons for further information
Exhibition dates
Company Profile